| Analyte | Calibration range (μmol/L) | LLMI (μmol/L) | Quality control (μmol/L) | Intra-day precisions (n=5) | Inter-day precisions(n=15) | Accuracy (n=5) |
| GABA | 0.15 | 1.97% | 2.28% | 103.83% | ||
| 0.05-10.00 | 0.02 | 3.00 | 1.22% | 1.65% | 101.06% | |
| 7.50 | 2.65% | 3.97% | 109.02% |
Fig 2 Effect of different mobile phase ratios on GABA response Chromatograms obtained with different mobile phase compositions. A: Mobile phase A1 (0.05% formic acid in water) and mobile phase B1 (90% acetonitrile in water) containing 5 mmol/L ammonium acetate; B: Mobile phase A1 (0.05% formic acid in water) and mobile phase B1 (90% acetonitrile in water) containing 0.10% formic acid and 5 mmol/L ammonium acetate; C: Mobile phase A1 (0.05% formic acid in water) and mobile phase B1 (90% acetonitrile in water) containing 0.175% formic acid and 5 mmol/L ammonium acetate.